Ubicación: Global + English
Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

The loss on drying (LOD) test is a fundamental analytical procedure used to determine the moisture and volatile matter content of chemical reagents by measuring the mass loss of a sample after drying under controlled conditions. This article provides a comprehensive technical overview of loss on drying testing for chemical reagents, covering testing principles, applicable sample types, standard test methods in accordance with GB/T 6284, JIS K 0068, and related international standards, calculation formulas, equipment requirements, and typical quality control applications.

Technical Overview

Loss on drying (LOD) is a gravimetric test that measures the percentage of mass lost when a chemical reagent sample is dried under specified conditions of temperature, pressure, and time. The mass loss typically represents free (surface) moisture, adsorbed water, and any volatile substances that evaporate under the selected drying conditions. Depending on the drying temperature, loosely bound water of crystallization may also be partially removed, so the test condition must always be matched to the thermal behavior of the specific reagent.

The test is governed by recognized standards, including GB/T 6284 (the general rule for the determination of loss on drying of chemical reagents), JIS K 0068 (test methods for loss on drying and residue on ignition of chemical products), and the general gravimetric principles described in ASTM and ISO methodologies. Pharmacopoeial monographs for reagent-grade and pharmaceutical-grade chemicals also specify loss-on-drying limits as a routine identity and purity criterion.

The result is calculated according to the following relationship:

Loss on drying (%) = (m1 − m2) / m1 × 100

where m1 is the mass of the sample before drying and m2 is the mass of the sample after drying to constant mass. Unlike Karl Fischer titration (e.g., ISO 760 or ASTM E203), which selectively quantifies water, loss on drying is a non-specific method that captures the total mass loss from all volatiles. It remains the preferred method for routine quality control because it is simple, robust, inexpensive, and directly traceable to mass measurement.

Testing Samples

Inorganic Salts and Hydrates

Common inorganic reagents such as sodium chloride, potassium nitrate, sodium carbonate, and sulfates are routinely tested for loss on drying. For crystalline hydrates (for example, copper(II) sulfate pentahydrate or sodium sulfate decahydrate), the drying temperature is deliberately selected to measure surface moisture without fully decomposing the crystal structure, unless the monograph specifically requires determination of total water of crystallization at an elevated temperature.

Organic Reagents and Fine Chemicals

Organic acids (such as citric acid and oxalic acid), sugars (such as glucose and lactose), buffers, and organic solvents with high boiling points are frequently tested. These materials often contain residual moisture from synthesis or crystallization steps, and the loss on drying value is used to verify compliance with the reagent grade specification.

Hygroscopic and Air-Sensitive Reagents

Strongly hygroscopic reagents, including various hydroxides, desiccants, and deliquescent salts, require rapid sample handling in a controlled atmosphere or glovebox to prevent moisture uptake during weighing. For such materials, accredited third-party laboratories apply accelerated transfer and weighing procedures to ensure that the measured loss reflects the sample itself rather than environmental contamination.

Pharmaceutical Intermediates, Catalysts, and Specialty Chemicals

Active pharmaceutical ingredient intermediates, catalyst precursors, adsorbents, and electronic-grade chemicals are tested because residual moisture affects reaction yield, catalyst activity, flowability, shelf life, and downstream process performance.

Testing Items

The scope of a loss on drying test program for chemical reagents typically includes the following determinations:

  • Loss on drying at a specified temperature (commonly 105 °C ± 2 °C) under atmospheric pressure
  • Loss on drying under reduced pressure (vacuum drying) for thermally sensitive reagents
  • Loss on drying at room temperature over a desiccant for volatile or low-melting substances
  • Residual volatile matter and solvent content expressed as mass percentage
  • Verification of drying to constant mass for gravimetric validity
  • Conversion of analytical assay results between the "as-is" basis and the "dried" basis

Testing Methods

Reference Oven-Drying Method (Constant-Mass Procedure)

The reference procedure, consistent with GB/T 6284 and equivalent international practices, is performed as follows:

  • Clean, dry, and pre-weigh a flat weighing bottle with its ground-glass lid, and confirm constant mass after a preliminary drying cycle in the oven and cooling in a desiccator.
  • Accurately weigh the specified quantity of sample, typically 1 g to 2 g or the amount stated in the product monograph, using an analytical balance with 0.1 mg readability.
  • Distribute the sample in the bottle as a uniform layer not exceeding 5 mm in thickness to ensure even heat penetration.
  • Dry the open bottle (with the lid alongside) in a forced-convection oven at the specified temperature, commonly 105 °C ± 2 °C, for approximately 2 hours.
  • Transfer the covered bottle to a desiccator charged with fresh silica gel or phosphorus pentoxide and allow it to cool to room temperature, typically 30 to 45 minutes.
  • Weigh accurately, then repeat drying cycles of about 1 hour until the difference between two consecutive weighings does not exceed 0.3 mg (or the criterion stated in the applicable standard), confirming constant mass.
  • Calculate the loss on drying as a mass percentage of the original sample.

Vacuum Drying Method

For reagents that decompose, oxidize, or melt at 105 °C, drying is performed in a vacuum oven at a reduced temperature (for example, 40 °C to 60 °C) and reduced pressure (commonly not greater than 5 kPa, or as specified). This accelerates moisture removal while protecting the chemical integrity of the sample.

Desiccator Method at Ambient Temperature

Substances that are heat-sensitive or contain volatile components may be dried at room temperature in a desiccator over an effective desiccant until constant mass is reached. This method is slow but gentle and is specified in certain reagent monographs.

Rapid Complementary Techniques

Infrared and halogen-lamp moisture analyzers, as well as thermogravimetric analysis (TGA), provide fast loss-on-drying estimates suitable for in-process screening. These rapid techniques must be validated against the reference oven method, with a documented correlation of drying temperature, time, and result equivalence before use in release testing.

Quality Control and Reporting

All determinations are performed at least in duplicate. The report should state the drying temperature, pressure, duration or constant-mass criterion, sample mass, individual and mean results, and the applicable acceptance limit. Calibrated balances, validated ovens with uniform temperature distribution, and active desiccants are prerequisites for reliable data in accredited third-party laboratories and in-house QC facilities alike.

Applications

Loss on drying testing for chemical reagents supports a wide range of quality and regulatory functions:

  • Incoming inspection of raw materials and reagent-grade chemicals against certified specifications
  • Final product release testing and issuance of certificates of analysis
  • Correction of assay and content results to a dry basis for accurate stoichiometric calculations
  • Long-term stability studies and accelerated storage condition evaluation
  • Verification of drying, packaging, and storage processes in reagent manufacturing
  • Qualification of moisture-sensitive reagents used in Karl Fischer titration, gravimetric analysis, and moisture-critical syntheses
  • Support for method validation, supplier audits, and compliance with GB, ISO, ASTM, EN, and JIS-based quality management systems

FAQ

What is the difference between loss on drying and water content?

Loss on drying measures the total mass lost under the drying conditions, which may include water plus volatile solvents and other volatiles. Water content determined by Karl Fischer titration is specific to water molecules only. The two values agree only when no other volatile substances are present in the reagent.

What does "constant mass" mean?

Constant mass is achieved when the difference between two consecutive weighings, after repeated drying and cooling cycles, does not exceed the specified tolerance, typically 0.3 mg. This confirms that the drying process is complete and the result is gravimetrically valid.

Why must samples be cooled in a desiccator before weighing?

A hot weighing bottle generates convection currents around the balance pan that cause inaccurate low readings, and exposure to ambient air allows reabsorption of moisture. Cooling in a desiccator isolates the sample from atmospheric humidity and ensures consistent, accurate weighing.

How is the drying temperature selected?

The temperature is specified by the product monograph or the applicable standard. A common default is 105 °C ± 2 °C for stable inorganic reagents; lower temperatures under vacuum are selected for heat-sensitive, meltable, or oxidizable substances to avoid thermal decomposition biasing the result.

Can the test be applied to reagents that sublime or decompose on heating?

Not directly. Sublimation or decomposition produces mass loss that is not attributable to moisture. For such reagents, an alternative drying condition (lower temperature, vacuum, or desiccator method) or a specific water determination method must be selected and justified.

How are hygroscopic samples handled?

Hygroscopic reagents are weighed rapidly in covered vessels, handled with gloves and tweezers to prevent contamination, and transferred between oven, desiccator, and balance with minimal air exposure. If necessary, weighing is performed in a dry inert-atmosphere enclosure.

What is the typical turnaround time for a loss on drying test?

A single determination typically requires 4 to 8 hours, including drying cycles, cooling periods, and confirmation of constant mass. Samples requiring prolonged ambient desiccator drying or repeat cycles may take longer; rapid moisture analyzers can deliver screening results within minutes when a validated correlation exists.