Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

This article provides a comprehensive technical analysis of CVCM mass change, a critical parameter for evaluating the outgassing properties of spacecraft materials. Understanding Collected Volatile Condensable Materials is essential for ensuring the reliability of optical instruments and thermal control surfaces in the vacuum of space.

CVCM Mass Change Analysis

Technical Overview

CVCM stands for Collected Volatile Condensable Materials. It refers to the fraction of outgassed molecules from a non-metallic material that condenses on a collector plate maintained at a specific temperature, typically 25 degrees Celsius, under vacuum conditions. The analysis of CVCM mass change is a fundamental aspect of space material screening, governed primarily by the ASTM E595 standard. When materials are exposed to the vacuum of space, they release volatile compounds. If these compounds condense on sensitive surfaces like lenses, mirrors, or solar cells, they can degrade mission performance. Therefore, quantifying this mass change is vital for contamination control.

Testing Samples

The selection and preparation of testing samples are critical for obtaining accurate CVCM data. The samples must represent the material as it will be used in the flight configuration.

Polymeric and Non-Metallic Materials

Most testing focuses on polymeric materials such as adhesives, potting compounds, films, and thermal control coatings. These materials typically possess higher levels of volatile constituents compared to metals. Samples are usually prepared in small, pre-weighed containers, ensuring a consistent surface area for outgassing.

Pre-Conditioning Requirements

Before testing, samples undergo a specific conditioning process to stabilize their initial mass. This typically involves drying the samples in a controlled environment to remove ambient moisture, which ensures that the subsequent mass change is strictly due to outgassing under vacuum rather than simple dehydration.

Testing Items

The CVCM analysis is part of a broader triad of measurements used to characterize material outgassing. The primary testing items include:

  • Total Mass Loss (TML): The percentage of mass lost by the sample after exposure to vacuum and heat.
  • Collected Volatile Condensable Materials (CVCM): The percentage of mass condensed on the collector plate relative to the initial sample mass.
  • Water Vapor Recovered (WVR): The amount of mass recovered by the sample after exposure to a humid environment following the vacuum test.

Testing Methods

The standard procedure for analyzing CVCM mass change follows a rigorous protocol designed to simulate the space environment while accelerating the outgassing process.

Test Setup and Conditions

The test utilizes a vacuum chamber maintained at a pressure below 1.0 x 10^-5 Torr. The samples are heated to 125 degrees Celsius to accelerate molecular kinetics. A collector plate, positioned near the sample, is maintained at 25 degrees Celsius to act as a cold trap for condensable species.

Mass Measurement Procedure

Initially, the sample and the collector plate are weighed with high precision using a microbalance. After a 24-hour exposure period in the vacuum chamber, the items are re-weighed. The CVCM value is calculated by dividing the mass gain of the collector plate by the initial mass of the sample. This mass change analysis reveals the contamination potential of the material.

Applications

Data derived from CVCM mass change analysis is utilized across various stages of spacecraft development and operation.

  • Material Selection: Engineers use CVCM data to reject materials that exceed the standard acceptance limit of 0.10%.
  • Contamination Analysis: For complex assemblies, CVCM data helps model the accumulation of molecular films on critical sensors over the mission lifetime.
  • Thermal Control Design: Designers must ensure that paints and coatings do not outgas heavily, which would alter their thermo-optical properties.

FAQ

Below are common questions regarding CVCM analysis and its implications for material science.

What is the standard acceptance criterion for CVCM?

According to NASA and ESA standards, a material is generally considered acceptable for spacecraft use if the CVCM value is below 0.10%. Materials exceeding this threshold pose a significant risk of contaminating optical and thermal surfaces.

How does temperature affect CVCM results?

Temperature directly influences the vapor pressure of contaminants. Heating the sample to 125 degrees Celsius simulates the worst-case scenario for outgassing, ensuring that materials passing this test will perform safely at lower operational temperatures in space.

Why is TML analyzed alongside CVCM?

TML indicates the total amount of material leaving the sample, while CVCM indicates how much of that material is likely to stick to surfaces. A material might have high TML but low CVCM, meaning it releases gas that does not condense, posing less of a contamination risk.